Computer Aided Process Improvement University of Florida
HYSYS Library

Operating Modes
Equipment Description
Simulation Basis
Continuous Distillation
Making A Run
Total Reflux
Batch Distillation
Acknowledgements
Further Readings

Computer Tools>HYSYS Plant>Distillation Tutorial Up

Modeling Continuous Steady State Distillation


Once you enter the simulation environment, you will be given a green screen where you can start building your model. 

Select The Model And Define Connecting Streams

For the 24-plate continuous distillation column case, follow these instructions:

  1. In the graphic area, select the distillation column by clicking on the icon shown by Figure 1 then clicking on the green screen. You will obtain a graph looking like Figure 2. You are ready to enter the different parameters for the column.
  2. Double click on the icon that looks like Figure 2 to enter the different specifications.
  3. First, you should enter the number of stages, which is 24 for the continuous column.  To do so click on the number of stages then go to the bar at the top of the screen then click to edit.
  4.  Select TOP DOWN to start numbering the stages from the top.
     
                                                       
    Figure 2
  5. Name your feed (we will name it here F). Assume the feed to enter at plate 10. (You can change that option later to adjust your output from the column).
  6. Select TOTAL REFLUX because we will only draw liquid from the condenser. Name your distillate stream (in this example D) and your bottom stream (in this example B). Also name your condenser and reboiler duties (in this example respectively Qc and Qr). Once done, click anywhere in the screen and the NEXT button will appear bold indicating we are ready to move to the next step. Click on it.
  7. In the laboratory we operate at atmospheric pressure. Therefore both the reboiler and the condenser will be at 1 atm. Enter the pressure by clicking on the appropriate box then on the bar up the screen. Enter 1 then in the small window next to the input bar, select by clicking on the arrow the right unit: atm. Repeat the same procedure for the other unit. We will leave the condenser pressure drop to be 0. Click on the NEXT button to continue.
  8. The next window displayed asks for estimated temperatures for the condenser and the reboiler. These temperatures are taken to be initial guesses when HYSYS performs iterations when finding variables. We will leave them blank and we will move on to the next screen.
  9. In the next window, we are asked to input the reflux ratio. In our example, it will be 3. We will not specify the liquid rate (otherwise our problem will be over specified). However, the liquid rate can be chosen as an independent variable, but one of the other variables we specified in the example would need to be dropped to avoid over specification. Click on DONE to continue. Then you will obtain a monitor window that will perform the calculations. Close that window to now enter the feed and other stream conditions.

Define The Feed Stream Parameters

In the laboratory as mentioned before, the feed is at room conditions (T=25oC, P=1atm). We will assume the concentration of the feed to be 50% ethanol, 50% isopropanol in mass fractions. (The concentration of the feed in the laboratory will be calculated prior to experiment). In our example we will also assume that we are feeding at a rate of 163 lb/hr. For the reboiler heat duty, we will assume a load of 120,000 BTU/hr.

  1. To input the parameters into the feed stream F, double click on the feed stream and you will obtain a window like Figure 3. In this window you will enter the characteristics necessary for the feed. In the Vapour/Phase Fraction box, enter 0 since the feed is purely liquid. Then click on the Temperature box to enter room temperature (25oC). Then click on pressure to enter the normal atmospheric pressure (1atm). Click on the Mass Flow box to input 163 lb/hr. The next step is to input the feed concentrations.
  2. To input the concentrations, click on COMP shown by the arrow on Figure 3. Click on EDIT. In the COMPOSITION BASIS section select Mass Fraction then enter 0.5 for each of the components. Then click OK. Close the window to go back to the main screen. You will notice that the feed stream changed color. This is because we have specified all the independent variables we care for on that stream and the dependant variables were automatically calculated by HYSYS.
     The next step is to enter the reboiler duty.  To do so, double click on the reboiler duty stream (Qr in our example).  When the screen appears, enter in the Heat Flow box the duty in BTU/hr (in our example 120,000 BTU/hr).  Click on CLOSE to go back to the main environment.
  3.  We need to inform HYSYS that the overhead and bottom products are liquid (no vapor).  To do so, double click on the top stream (in our case D) then input 0 in Vapour/Phase Fraction box, then click on CLOSE.  Repeat the same procedure for the bottom stream (B in our case).

 We are now ready to begin performing column calculations since all the variables have been specified.

Figure 3

Run A Column Simulation

We are now ready to run a column simulation to obtain all the dependent variables such as the overhead and bottom distillate conditions, the condenser load and so on. 

UF Virtual Tour Sign Up For A Campus Tour
Back Home Up Next
Privacy Policy  |  Disclaimer

Copyright ©1996-2001 Chemical Engineering DepartmentUniversity of Florida, Gainesville, FL 32611; (352) 392-0881.
This page was last updated Sunday, August 19, 2001 01:14:18 AM